PT-2015-1621

CVE-2015-1011

·

تاريخ النشر

2015-07-06

·

تاريخ التحديث

2015-07-08

الوصف من

CVSS 2.0
5.0
متوسطة
المقاييسAV:N/AC:L/Au:N/C:P/I:N/A:N
Name of the Vulnerable Software and Affected Versions Hospira LifeCare PCA Infusion System versions prior to 7.0
Description The issue is related to the lack of protection for service data in the device's software, which can be exploited by a remote attacker to gain access to the device. Hardcoded credentials in the system make it easier for attackers to obtain access.
Recommendations For versions prior to 7.0, update to version 7.0 or later to resolve the issue. As a temporary workaround, consider restricting access to the device to minimize the risk of exploitation.