PT-2015-3803
CVE-2014-5406
·
تاريخ النشر
2015-07-06
·
تاريخ التحديث
2015-07-08
الوصف من
CVSS 2.0
7.6
عاليةالمقاييسAV:N/AC:H/Au:N/C:C/I:C/A:C
Name of the Vulnerable Software and Affected Versions
Hospira LifeCare PCA Infusion System versions prior to 7.0
Description
The issue concerns the lack of validation of network traffic related to certain system updates or changes, allowing remote attackers to modify settings or medication data. This can be achieved via packets on various ports, including TELNET, HTTP, HTTPS, or UPNP.
Recommendations
For versions prior to 7.0, update to version 7.0 or later to resolve the issue.