PT-2019-13406

CVE-2019-13543

·

تاريخ النشر

2019-11-08

·

تاريخ التحديث

2019-11-13

الوصف من

CVSS 3.1
7.5
عالية
المقاييسAV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Name of the Vulnerable Software and Affected Versions Medtronic Valleylab Exchange Client version 3.4 and below Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below
Description The issue concerns the use of multiple sets of hard-coded credentials in the affected software. If these credentials are discovered, they can be used to read files on the device.
Recommendations For Medtronic Valleylab Exchange Client version 3.4 and below, update to a version above 3.4 to resolve the issue. For Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, update to a version above 4.0.0 to resolve the issue. For Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below, update to a version above 1.1.0 to resolve the issue.