PT-2019-7181
CVE-2015-3952
·
تاريخ النشر
2019-03-25
·
تاريخ التحديث
2019-10-09
الوصف من
CVSS 2.0
5.0
متوسطةالمقاييسAV:N/AC:L/Au:N/C:P/I:N/A:N
Name of the Vulnerable Software and Affected Versions
Hospira Plum A+ Infusion System versions 13.4 and prior
Hospira Plum A+3 Infusion System versions 13.6 and prior
Hospira Symbiq Infusion System versions 3.13 and prior
Description
The issue concerns the storage of wireless keys in plain text. To mitigate the risk, it is recommended to close Port 20/FTP and Port 23/TELNET on the affected devices.
Recommendations
For Hospira Plum A+ Infusion System versions 13.4 and prior, close Port 20/FTP and Port 23/TELNET.
For Hospira Plum A+3 Infusion System versions 13.6 and prior, close Port 20/FTP and Port 23/TELNET.
For Hospira Symbiq Infusion System versions 3.13 and prior, close Port 20/FTP and Port 23/TELNET.