PT-2019-7183
CVE-2015-3954
·
تاريخ النشر
2019-03-25
·
تاريخ التحديث
2019-10-09
الوصف من
CVSS 2.0
10
عاليةالمقاييسAV:N/AC:L/Au:N/C:C/I:C/A:C
Name of the Vulnerable Software and Affected Versions
Hospira Plum A+ Infusion System versions 13.4 and prior
Hospira Plum A+3 Infusion System versions 13.6 and prior
Hospira Symbiq Infusion System version 3.13 and prior
Description
The issue allows unauthenticated users to gain root privileges on the affected devices via Port 23/TELNET by default. This could enable an unauthorized user to issue commands to the pump.
Recommendations
For Hospira Plum A+ Infusion System versions 13.4 and prior, close Port 23/TELNET on the affected devices.
For Hospira Plum A+3 Infusion System versions 13.6 and prior, close Port 23/TELNET on the affected devices.
For Hospira Symbiq Infusion System version 3.13 and prior, close Port 23/TELNET on the affected devices.
وجدت خطأً في الوصف؟ لديك ما تضيفه؟ لا تتردد في مراسلتنا 👾