PT-2019-7184

CVE-2015-3956

·

تاريخ النشر

2019-03-25

·

تاريخ التحديث

2019-10-09

الوصف من

CVSS 2.0
10
عالية
المقاييسAV:N/AC:L/Au:N/C:C/I:C/A:C
Name of the Vulnerable Software and Affected Versions Hospira Plum A+ Infusion System versions 13.4 and prior Hospira Plum A+3 Infusion System versions 13.6 and prior Hospira Symbiq Infusion System version 3.13 and prior
Description The issue allows unauthorized devices on the host network to send drug libraries, firmware updates, pump commands, and make configuration changes to the affected infusion systems without authentication.
Recommendations For Hospira Plum A+ Infusion System versions 13.4 and prior, close Port 20/FTP and Port 23/TELNET on the affected devices. For Hospira Plum A+3 Infusion System versions 13.6 and prior, close Port 20/FTP and Port 23/TELNET on the affected devices. For Hospira Symbiq Infusion System version 3.13 and prior, close Port 20/FTP and Port 23/TELNET on the affected devices.
قابلية الاستغلال
إصلاح
التأثيرات
المورّدون
المنتجات المتأثرة
المعرّفات ذات الصلة
وجدت خطأً في الوصف؟ لديك ما تضيفه؟ لا تتردد في مراسلتنا 👾