Name of the Vulnerable Software and Affected Versions Baxter PrismaFlex all versions PrisMax versions prior to 3.x
Description The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.
Recommendations For Baxter PrismaFlex all versions, consider changing the hard-coded service password to a unique and secure password. For PrisMax versions prior to 3.x, update to version 3.x or later to remove the hard-coded service password. As a temporary workaround, consider restricting access to the device settings and calibration settings to minimize the risk of exploitation.