PT-2020-13004
CVE-2020-12037
·
تاريخ النشر
2020-06-29
·
تاريخ التحديث
2020-07-14
الوصف من
CVSS 3.1
7.5
عاليةالمقاييسAV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Name of the Vulnerable Software and Affected Versions
Baxter PrismaFlex all versions
PrisMax versions prior to 3.x
Description
The affected devices do not implement data-in-transit encryption, such as TLS/SSL, when sending treatment data to a Patient Data Management System (PDMS) or an Electronic Medical Record (EMR) system. This allows an attacker to observe sensitive data sent from the device.
Recommendations
For Baxter PrismaFlex all versions: Consider configuring the device to use data-in-transit encryption, such as TLS/SSL, when sending treatment data to a PDMS or EMR system.
For PrisMax versions prior to 3.x: Update to version 3.x or later, which is expected to include data-in-transit encryption, such as TLS/SSL, when sending treatment data to a PDMS or EMR system. As a temporary workaround, consider restricting access to the device's communication interface to minimize the risk of exploitation.
وجدت خطأً في الوصف؟ لديك ما تضيفه؟ لا تتردد في مراسلتنا 👾