PT-2022-8700

CVE-2020-25154

·

تاريخ النشر

2022-04-14

·

تاريخ التحديث

2022-04-21

الوصف من

CVSS 3.1
6.1
متوسطة
المقاييسAV:N/AC:L/PR:N/UI:R/S:C/C:L/I:L/A:N
Name of the Vulnerable Software and Affected Versions B. Braun Melsungen AG SpaceCom device versions L81/U61 and earlier B. Braun Melsungen AG Data module compactplus versions A10 and A11
Description The issue is related to an open redirect vulnerability in the administrative interface. This vulnerability allows attackers to redirect users to malicious websites.
Recommendations For B. Braun Melsungen AG SpaceCom device versions L81/U61 and earlier, consider restricting access to the administrative interface until a fix is available. For B. Braun Melsungen AG Data module compactplus versions A10 and A11, avoid using the administrative interface for sensitive operations until the issue is resolved. At the moment, there is no information about a newer version that contains a fix for this vulnerability.
التأثيرات
المورّدون