PT-2018-16957
CVE-2018-5438
·
Publicado
2018-03-20
·
Atualizado
2018-04-20
Descrição por
CVSS 3.1
6.3
MédiaMétricasAV:L/AC:H/PR:L/UI:N/S:U/C:H/I:H/A:N
Name of the Vulnerable Software and Affected Versions
Philips ISCV versions prior to 2.3.0
Description
The issue is related to insufficient session expiration, allowing an attacker to reuse a previously logged-in user's session. This occurs when the ISCV application is used with an Electronic Medical Record (EMR) system in KIOSK mode for multiple users, and Windows authentication is in use. An attacker could potentially gain unauthorized access to patient health information and modify it.
Recommendations
For versions prior to 2.3.0, update to version 2.3.0 or later to resolve the issue. As a temporary workaround, consider restricting access to the ISCV application when used in KIOSK mode with Windows authentication to minimize the risk of exploitation.