PT-2018-5443
CVE-2017-12712
·
Publicado
2018-04-25
·
Atualizado
2019-10-09
Descrição por
CVSS 3.1
8.8
AltaMétricasAV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Name of the Vulnerable Software and Affected Versions:
Abbott Laboratories pacemakers versions prior to Aug 28, 2017
Description:
The authentication algorithm in the pacemakers can be compromised or bypassed, allowing a nearby attacker to issue unauthorized commands via RF communications. This involves an authentication key and time stamp.
Recommendations:
For pacemakers manufactured prior to Aug 28, 2017, apply the firmware update developed by Abbott to help mitigate the identified issue.
Explorabilidade
Correção
Impactos
Fornecedores
Produtos afetados
Encontrou algum problema na descrição? Tem algo a acrescentar? Fique à vontade para nos escrever 👾