PT-2018-5443

CVE-2017-12712

·

Publicado

2018-04-25

·

Atualizado

2019-10-09

Descrição por

CVSS 3.1
8.8
Alta
MétricasAV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Name of the Vulnerable Software and Affected Versions: Abbott Laboratories pacemakers versions prior to Aug 28, 2017
Description: The authentication algorithm in the pacemakers can be compromised or bypassed, allowing a nearby attacker to issue unauthorized commands via RF communications. This involves an authentication key and time stamp.
Recommendations: For pacemakers manufactured prior to Aug 28, 2017, apply the firmware update developed by Abbott to help mitigate the identified issue.