PT-2019-7128
CVE-2014-5432
·
Publicado
2019-03-26
·
Atualizado
2019-10-09
Descrição por
CVSS 2.0
7.5
AltaMétricasAV:N/AC:L/Au:N/C:P/I:P/A:P
Name of the Vulnerable Software and Affected Versions
Baxter SIGMA Spectrum Infusion System version 6.05
Description
The issue allows remote access via Port 22/SSH without authentication, enabling a remote attacker to make unauthorized configuration changes to the wireless battery module (WBM) and potentially access account credentials and shared keys. It is noted that this vulnerability does not allow control of the SIGMA Spectrum infusion pump from the WBM.
Recommendations
For Baxter SIGMA Spectrum Infusion System version 6.05, update to Version 8, which incorporates necessary hardware and software changes to address the issue.
Explorabilidade
Correção
Impactos
Fornecedores
Produtos afetados
Encontrou algum problema na descrição? Tem algo a acrescentar? Fique à vontade para nos escrever 👾