PT-2019-7129

CVE-2014-5433

·

Publicado

2019-03-26

·

Atualizado

2019-10-09

Descrição por

CVSS 2.0
7.5
Alta
MétricasAV:N/AC:L/Au:N/C:P/I:P/A:P
Name of the Vulnerable Software and Affected Versions Baxter SIGMA Spectrum Infusion System version 6.05
Description An unauthenticated remote attacker may be able to execute commands to view wireless account credentials stored in cleartext, potentially allowing access to the host network.
Recommendations For Baxter SIGMA Spectrum Infusion System version 6.05, update to version 8, which incorporates necessary hardware and software changes to address the issue.