PT-2019-7184
CVE-2015-3956
·
Publicado
2019-03-25
·
Atualizado
2019-10-09
Descrição por
CVSS 2.0
10
AltaMétricasAV:N/AC:L/Au:N/C:C/I:C/A:C
Name of the Vulnerable Software and Affected Versions
Hospira Plum A+ Infusion System versions 13.4 and prior
Hospira Plum A+3 Infusion System versions 13.6 and prior
Hospira Symbiq Infusion System version 3.13 and prior
Description
The issue allows unauthorized devices on the host network to send drug libraries, firmware updates, pump commands, and make configuration changes to the affected infusion systems without authentication.
Recommendations
For Hospira Plum A+ Infusion System versions 13.4 and prior, close Port 20/FTP and Port 23/TELNET on the affected devices.
For Hospira Plum A+3 Infusion System versions 13.6 and prior, close Port 20/FTP and Port 23/TELNET on the affected devices.
For Hospira Symbiq Infusion System version 3.13 and prior, close Port 20/FTP and Port 23/TELNET on the affected devices.
Explorabilidade
Correção
Impactos
Fornecedores
Produtos afetados