PT-2019-7868

CVE-2017-10724

·

Publicado

2019-06-17

·

Atualizado

2019-06-20

Descrição por

CVSS 2.0
6.5
Média
MétricasAV:N/AC:L/Au:S/C:P/I:P/A:P
Name of the Vulnerable Software and Affected Versions Shekar Endoscope (affected versions not specified)
Description A memory corruption issue was discovered in the Shekar Endoscope device, allowing an attacker connected to the device's Wi-Fi SSID to execute remote code on the device. This device is used in various industrial systems, car garages, and medical clinics to access areas difficult for humans to reach. A breach of this system can allow an attacker to access the video feed and pictures viewed by the user and potentially gain a foothold in air-gapped networks, especially in nation-critical infrastructure and industries. The firmware contains a binary uvc stream, which is a UDP daemon responsible for handling UDP requests. The client application sends a UDP request to change the Wi-Fi name, which is handled by the control Dev thread function. This function calls setwifipassword, which uses a memcpy function with the length of the payload obtained by the strlen function, allowing an attacker to overflow the function and control the $PC value.
Recommendations At the moment, there is no information about a newer version that contains a fix for this vulnerability.
Explorabilidade
Exploit
Impactos
Pesquisadores
Fornecedores
Produtos afetados
Identificadores relacionados
Encontrou algum problema na descrição? Tem algo a acrescentar? Fique à vontade para nos escrever 👾