PT-2023-20366 · Medtronic · Medtronic Pelvic Health Clinician Apps
CVE-2023-25931
·
Publicado
2023-03-01
·
Atualizado
2023-03-09
CVSS v3.1
6.8
Média
| Vetor | AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
Name of the Vulnerable Software and Affected Versions
Medtronic Pelvic Health clinician apps (affected versions not specified)
Description
The issue is related to a password vulnerability in the Medtronic Pelvic Health clinician apps installed on the Smart Programmer mobile device. This vulnerability could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. However, changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.
Recommendations
At the moment, there is no information about a newer version that contains a fix for this vulnerability.
Improper Authentication