PT-2023-30398 · Abbott · Id Now

CVE-2023-47262

·

Publicado

2023-11-14

·

Atualizado

2023-12-13

CVSS v3.1

5.2

Média

VetorAV:P/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:L
Name of the Vulnerable Software and Affected Versions Abbott ID NOW versions prior to 7.1
Description The issue allows modification of the startup process and device configurations via physical access to an internal serial port. Direct physical access is required to exploit this issue. Settings can be modified through this access.
Recommendations For versions prior to 7.1, as a temporary workaround, consider restricting physical access to the internal serial port until a patch is available. At the moment, there is no information about a newer version that contains a fix for this vulnerability.
Encontrou algum problema na descrição? Tem algo a acrescentar? Fique à vontade para nos escrever 👾

Identificadores relacionados

CVE-2023-47262

Produtos afetados

Id Now