PT-2015-3803
CVE-2014-5406
·
Published
2015-07-06
·
Updated
2015-07-08
Description by
CVSS 2.0
7.6
HighMetricsAV:N/AC:H/Au:N/C:C/I:C/A:C
Name of the Vulnerable Software and Affected Versions
Hospira LifeCare PCA Infusion System versions prior to 7.0
Description
The issue concerns the lack of validation of network traffic related to certain system updates or changes, allowing remote attackers to modify settings or medication data. This can be achieved via packets on various ports, including TELNET, HTTP, HTTPS, or UPNP.
Recommendations
For versions prior to 7.0, update to version 7.0 or later to resolve the issue.
Found an issue in the description? Have something to add? Feel free to write us 👾