PT-2018-5443
CVE-2017-12712
·
Published
2018-04-25
·
Updated
2019-10-09
Description by
CVSS 3.1
8.8
HighMetricsAV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Name of the Vulnerable Software and Affected Versions:
Abbott Laboratories pacemakers versions prior to Aug 28, 2017
Description:
The authentication algorithm in the pacemakers can be compromised or bypassed, allowing a nearby attacker to issue unauthorized commands via RF communications. This involves an authentication key and time stamp.
Recommendations:
For pacemakers manufactured prior to Aug 28, 2017, apply the firmware update developed by Abbott to help mitigate the identified issue.
Found an issue in the description? Have something to add? Feel free to write us 👾