PT-2018-5444

CVE-2017-12714

·

Published

2018-04-25

·

Updated

2019-10-09

Description by

CVSS 3.1
6.5
Medium
MetricsAV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Name of the Vulnerable Software and Affected Versions: Abbott Laboratories pacemakers (affected versions not specified), however, it is mentioned that the affected devices are those manufactured prior to Aug 28, 2017.
Description: The issue concerns pacemakers that do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received. This may allow a nearby attacker to repeatedly send commands, potentially reducing pacemaker battery life.
Recommendations: For pacemakers manufactured prior to Aug 28, 2017, apply the firmware update developed by Abbott to help mitigate the issue.
Exploitability
Fix
Vendors
Affected products
Related identifiers