PT-2019-7128

CVE-2014-5432

·

Published

2019-03-26

·

Updated

2019-10-09

Description by

CVSS 2.0
7.5
High
MetricsAV:N/AC:L/Au:N/C:P/I:P/A:P
Name of the Vulnerable Software and Affected Versions Baxter SIGMA Spectrum Infusion System version 6.05
Description The issue allows remote access via Port 22/SSH without authentication, enabling a remote attacker to make unauthorized configuration changes to the wireless battery module (WBM) and potentially access account credentials and shared keys. It is noted that this vulnerability does not allow control of the SIGMA Spectrum infusion pump from the WBM.
Recommendations For Baxter SIGMA Spectrum Infusion System version 6.05, update to Version 8, which incorporates necessary hardware and software changes to address the issue.
Exploitability
Fix
Impacts
Vendors
Affected products
Related identifiers
Found an issue in the description? Have something to add? Feel free to write us 👾