PT-2019-7182

CVE-2015-3953

·

Published

2019-03-25

·

Updated

2019-10-09

Description by

CVSS 2.0
10
High
MetricsAV:N/AC:L/Au:N/C:C/I:C/A:C
Name of the Vulnerable Software and Affected Versions Hospira Plum A+ Infusion System versions 13.4 and prior Hospira Plum A+3 Infusion System versions 13.6 and prior Hospira Symbiq Infusion System version 3.13 and prior
Description The issue allows access to the system through hard-coded accounts. To mitigate this, closing specific ports is recommended. There are no reported real-world incidents or estimated numbers of affected devices provided.
Recommendations For Hospira Plum A+ Infusion System versions 13.4 and prior, close Port 20/FTP and Port 23/TELNET. For Hospira Plum A+3 Infusion System versions 13.6 and prior, close Port 20/FTP and Port 23/TELNET. For Hospira Symbiq Infusion System version 3.13 and prior, close Port 20/FTP and Port 23/TELNET.
Exploitability
Fix
Impacts
Vendors
Affected products
Related identifiers
Found an issue in the description? Have something to add? Feel free to write us 👾