PT-2020-12227

CVE-2020-10598

·

Published

2020-04-01

·

Updated

2021-09-14

Description by

CVSS 3.1
6.1
Medium
MetricsAV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
Name of the Vulnerable Software and Affected Versions BD Pyxis MedStation ES System version 1.6.1 BD Pyxis Anesthesia (PAS) ES System version 1.6.1
Description A restricted desktop environment escape issue exists in the kiosk mode functionality of affected devices. Specially crafted inputs could allow a user to escape the restricted environment, resulting in access to sensitive data.
Recommendations For BD Pyxis MedStation ES System version 1.6.1, consider disabling the kiosk mode functionality until a patch is available. For BD Pyxis Anesthesia (PAS) ES System version 1.6.1, consider disabling the kiosk mode functionality until a patch is available.