PT-2020-13014

CVE-2020-12048

·

Published

2020-06-29

·

Updated

2020-07-16

Description by

CVSS 3.1
7.5
High
MetricsAV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Name of the Vulnerable Software and Affected Versions Phoenix Hemodialysis Delivery System SW versions 3.36 through 3.40
Description The issue concerns the lack of data-in-transit encryption, such as TLS/SSL, when transmitting sensitive treatment and prescription data between the Phoenix system and the Exalis dialysis data management tool over the network. This could allow an attacker with network access to observe the sensitive data being transmitted.
Recommendations For Phoenix Hemodialysis Delivery System SW versions 3.36 through 3.40, consider implementing encryption for data transmitted between the Phoenix system and the Exalis tool to protect sensitive information. As a temporary workaround, restrict network access to authorized personnel only to minimize the risk of exploitation.
Exploitability
Fix
Impacts
Vendors
Affected products
Related identifiers
Found an issue in the description? Have something to add? Feel free to write us 👾