PT-2020-14453

CVE-2020-15486

·

Published

2020-08-26

·

Updated

2021-07-21

Description by

CVSS 3.1
6.5
Medium
MetricsAV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Name of the Vulnerable Software and Affected Versions Dr Trust ECG Pen version 2.00.08
Description An issue allows attackers to access the GATT server of the device without requiring pairing or security. This enables attackers to sniff data being broadcasted during measurements and extract saved data over a Bluetooth connection. Additionally, attackers can launch a man-in-the-middle attack against data integrity.
Recommendations For Dr Trust ECG Pen version 2.00.08, consider disabling Bluetooth LE support until a secure pairing or security mechanism is implemented to prevent unauthorized access. Restrict access to the GATT server to minimize the risk of data sniffing and extraction. Avoid using the device for sensitive measurements until the issue is resolved.
Exploitability
Fix
Researchers
Vendors
Affected products
Related identifiers
Found an issue in the description? Have something to add? Feel free to write us 👾