PT-2020-16003
CVE-2020-25187
·
Published
2020-12-14
·
Updated
2020-12-15
Description by
CVSS 3.1
8.8
HighMetricsAV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H
Name of the Vulnerable Software and Affected Versions
Medtronic MyCareLink Smart 25000 all versions
Description
The issue allows an attacker who gains authentication to run a debug command, which is sent to the reader, causing a heap overflow in the MCL Smart Reader stack. This heap overflow enables the attacker to remotely execute code on the MCL Smart Reader, potentially leading to control of the device.
Recommendations
For Medtronic MyCareLink Smart 25000 all versions, as a temporary workaround, consider restricting access to the debug command until a patch is available. Additionally, restrict access to the MCL Smart Reader to minimize the risk of exploitation. At the moment, there is no information about a newer version that contains a fix for this vulnerability.