PT-2020-16666

CVE-2020-27252

·

Published

2020-12-14

·

Updated

2020-12-15

Description by

CVSS 3.1
8.8
High
MetricsAV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H
Name of the Vulnerable Software and Affected Versions Medtronic MyCareLink Smart 25000 all versions
Description The issue concerns a race condition in the MCL Smart Patient Reader software update system, allowing unsigned firmware to be uploaded and executed. This could enable an attacker to remotely execute code on the device, potentially leading to device control.
Recommendations For Medtronic MyCareLink Smart 25000 all versions: At the moment, there is no information about a newer version that contains a fix for this vulnerability.
Impacts
Vendors
Affected products
Related identifiers
Found an issue in the description? Have something to add? Feel free to write us 👾