PT-2020-19309

CVE-2020-6961

·

Published

2020-01-24

·

Updated

2020-03-17

Description by

CVSS 3.1
10
Critical
MetricsAV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Name of the Vulnerable Software and Affected Versions ApexPro Telemetry Server versions 4.2 and prior CARESCAPE Telemetry Server versions 4.3 and prior Clinical Information Center (CIC) versions 4.X and 5.X CARESCAPE Central Station (CSCS) versions 1.X
Description A vulnerability exists in the affected products that could allow an attacker to obtain access to the SSH private key in configuration files.
Recommendations For ApexPro Telemetry Server versions 4.2 and prior, update to a version later than 4.2 to resolve the issue. For CARESCAPE Telemetry Server versions 4.3 and prior, update to a version later than 4.3 to resolve the issue. For Clinical Information Center (CIC) versions 4.X and 5.X, restrict access to configuration files containing SSH private keys until a patch is available. For CARESCAPE Central Station (CSCS) versions 1.X, consider disabling SSH access temporarily until a fix is provided.
Exploitability
Fix
Impacts
Vendors
Affected products
Related identifiers
Found an issue in the description? Have something to add? Feel free to write us 👾