PT-2026-45627 · Dräger · Infinity Explorer C700
CVE-2019-25718
·
Published
2026-06-01
·
Updated
2026-07-22
CVSS v3.1
8.4
High
| Vector | AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
Name of the Vulnerable Software and Affected Versions
Dräger Infinity Explorer C700 (affected versions not specified)
Description
A privilege escalation issue allows attackers to break out of kiosk mode—a restricted user interface that limits access to specific applications—and access the underlying operating system through a specific dialog interaction. This kiosk escape can be used to take control of the operating system, potentially causing the device to display incorrect information or no information from the connected Delta Family patient monitor.
Recommendations
At the moment, there is no information about a newer version that contains a fix for this vulnerability.
LPE
UI Misrepresentation of Critical Information
Found an issue in the description? Have something to add? Feel free to write us 👾
Weakness Enumeration
Related Identifiers
Affected Products
Infinity Explorer C700