PT-2026-67126 · Thermo Fisher Applied Biosystems · Genetic Analyzers

·

CVE-2026-17583

·

Published

2026-08-03

·

Updated

2026-08-06

CVSS v3.1

8.4

High

VectorAV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Name of the Vulnerable Software and Affected Versions Thermo Fisher Applied Biosystems Genetic Analyzers (affected versions not specified)
Description Output files with .fsa or .hid extensions can be edited, allowing an attacker to tamper with DNA data. This could lead to inaccurate DNA test outcomes, as demonstrated by a researcher who successfully combined two different profiles into a single file without triggering any warnings.
Recommendations At the moment, there is no information about a newer version that contains a fix for this vulnerability.
Found an issue in the description? Have something to add? Feel free to write us 👾

Weakness Enumeration

Related Identifiers

CVE-2026-17583

Affected Products

Genetic Analyzers