PT-2023-30615

CVE-2023-47800

·

Published

2023-11-09

·

Updated

2023-11-23

Description by

CVSS 3.1
9.8
Critical
MetricsAV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Name of the Vulnerable Software and Affected Versions Natus NeuroWorks and SleepWorks versions prior to 8.4 GMA3
Description The issue arises from the use of a default password xltek for the Microsoft SQL Server service sa account in Natus NeuroWorks and SleepWorks. This allows a threat actor to perform remote code execution, data exfiltration, or other malicious actions such as tampering with data or destroying/disrupting MSSQL services.
Recommendations For versions prior to 8.4 GMA3, update to version 8.4 GMA3 or later to resolve the issue. As a temporary workaround, consider changing the default password xltek for the Microsoft SQL Server service sa account to a strong, unique password. Restrict access to the Microsoft SQL Server service to minimize the risk of exploitation.
Exploitability
Exploit
Fix
Impacts
Researchers
Vendors
Affected products
Related identifiers
Found an issue in the description? Have something to add? Feel free to write us 👾